Home Precision Environment Test Systems AFC 132 Filter Media & Small Filter Element Test System
AFC 132 Filter Media & Small Filter Element Test System product image thumbnail
AFC 132 Filter Media & Small Filter Element Test System workflow diagram thumbnail
AFC 132 Filter Media & Small Filter Element Test System application overview thumbnail
AFC 132 Filter Media & Small Filter Element Test System application overview thumbnail
Topas GmbH Filter Media & Element Test Systems

AFC 132 Filter Media & Small Filter Element Test System

Flexible test system for filter media and small filter elements, supporting pressure-drop, gravimetric-efficiency, fractional-efficiency and dust-holding measurements across multiple air-filter standards.

Test flow
0.4–40 or 2–64 m³/h
Specimen size
Diameter ≤ 150 mm
Particle sizing
0.2–10 µm typical
Aerosols
DEHS / oil / NaCl / PSL
Filter media Small elements Variable flow Fractional efficiency Multi-standard
ISO 16890-2 ISO 16890-3 EN 779 ASHRAE 52.2 DIN 71460-1 ISO 5011 ISO 29463-3

Model Overview

Model System role Primary measurements Standards / application fit
AFC 132 Filter-media and small-element test system Pressure drop, gravimetric efficiency, fractional efficiency and dust holding capacity General, cabin, intake and high-efficiency filter development and quality assurance

Key Features

Multi-standard platform

Supports test concepts for general air filters, cabin filters, intake filters and high-efficiency media.

Liquid and solid aerosols

Can generate or use DEHS, paraffin oil, NaCl, PSL and customer-specific test aerosols.

Variable volume-flow range

Alternative measuring sections support different face velocities and specimen types.

Single-instrument sample switching

Upstream, downstream and purge sampling can be switched through a controlled sampling unit.

Flexible system expansion

Modular Topas components allow later changes in aerosol generation, dilution and particle measurement.

AFC132Win software

Integrated control and data acquisition support repeatable measurement sequences and reports.

Technical Specifications

Parameter Unit Value / Description
TEST RANGE
Test flow rate m³/h 0.4–40 or 2–64
Specimen size Diameter ≤ 150 mm
Typical particle-size range µm 0.2–10 with LAP 340; optional wider-range instrumentation
MEDIA & UTILITIES
Test aerosols DEHS, paraffin oil, NaCl, PSL and customer-specific aerosols
Compressed-air supply bar 6
Power supply 400 VAC, 16 A
Dimensions mm 2,100 × 1,400 × 850

Values shown are based on the standard system configuration. Confirm the final test method, specimen fixtures, instrumentation, utilities and calibration scope before order placement.

Applications

Ventilation-filter media

ISO 16890 and related pressure-drop and efficiency measurements.

High-efficiency media

Filter-media and small-element testing in support of ISO 29463 development.

Cabin and intake filters

Small-scale testing aligned with cabin-air and engine-intake filter methods.

Incoming and final inspection

Quality assurance for flat media and small elements before further processing.